1.
Braz. dent. j ; 34(2): 21-26, Mar.-Apr. 2023. tab, graf
Article in English | LILACS-Express | LILACS, BBO | ID: biblio-1439577

ABSTRACT

Abstract The present study evaluated the actual rotational speed of three different endodontic motors compared to the values provided by the manufacturers. A total of three endodontic motors (X-Smart Plus, VDW.Silver, and iRoot) were tested at 400 rpm and 800 rpm and 2 N/cm2 torque. The kinematics of the devices was recorded by using a custom angle-measuring disc with a 50-mm diameter attached to the handpiece provided by the manufacturer, whereas their movement was captured by a high-speed camera at 2,400 frames per second, 800 x 800 pixel-resolution and distance of 0.3 m from the target object. Statistical analysis was performed at a significance level of 5%. At 400 rpm, the iRoot motor had a value of 17.94 rpm above that indicated by the manufacturer, which was significantly different from those of X-Smart Plus (5.20 rpm below that indicated by the manufacturer) and VDW.Silver (0.62 rpm above that indicated by the manufacturer) motors (P < 0.05). At 800 rpm, the iRoot motor had a value of 51.34 rpm below that indicated by the manufacturer, whereas the X-Smart Plus motor had a value of 13.00 rpm below that indicated by the manufacturer (P > 0.05). The VDW.Silver motor statistically differed from the iRoot and X-Smart Plus ones, showing a value of 1.68 rpm above that indicated by the manufacturer. In conclusion, the X-Smart Plus, VDW.Silver, and iRoot motors showed lower variations in the rotational speed values compared to those reported by their manufacturers. The endodontic motors presented different behaviors between them, with the VDW.Silver motor presents the most accurate values and the iRoot presents the most divergent values.


Resumo O presente estudo avaliou a real velocidade rotacional de três diferentes motores endodônticos comparados aos valores fornecidos pelos fabricantes. Um total de três motores endodônticos (X-Smart Plus, VDW.Silver e iRoot) foram testados a 400 rpm e 800 rpm e torque de 2 N/cm2. A cinemática dos dispositivos foi verificada utilizando-se um disco customizado de 50 mm de diâmetro acoplado ao contra ângulo fornecido pelo fabricante e o movimento foi capturado através de uma câmera ajustada em 2400 quadros por segundo com resolução de 800 x 800 pixels distante 0,3 m do objeto alvo. A análise estatística foi realizada com nível de significância de 5%. A 400 rpm, o motor iRoot apresentou 17,94 rpm acima do indicado pelo fabricante, sendo o valor significativamente diferente do X-Smart Plus (5,20 rpm abaixo do indicado pelo fabricante) e VDW.Silver (0.62 rpm acima do indicado pelo fabricante) (P > 0,05). A 800 rpm, iRoot apresentou 51.34 rpm abaixo do indicado pelo fabricante e X-Smart Plus 13,00 rpm abaixo do indicado pelo fabricante (P > 0,05). VDW.Silver foi estatisticamente diferente de iRoot e X-Smart Plus, apresentando 1,68 rpm acima do indicado pelo fabricante. Concluiu-se que os motores X-Smart Plus, VDW.Silver e iRoot apresentaram baixa variação na velocidade rotacional quando comparada ao fornecido pelo fabricante. Os motores endodônticos apresentaram diferentes comportamentos entre eles, sendo que o motor VDW.Silver apresentou valores mais acurados enquanto o iRoot foi o mais divergente.

2.
Braz. dent. j ; 26(6): 624-629, Nov.-Dec. 2015. graf
Article in English | LILACS | ID: lil-769560

ABSTRACT

The aim of this study was to compare the percentage of non-instrumented area of root canals prepared with different enlargements using single-file reciprocating systems (Reciproc and WaveOne) and a conventional multi-file rotary (BioRaCe) system by micro-computed tomographic analysis. Thirty mesial roots of mandibular molars with moderate curvature (10° to 20°) presenting a type II Vertucci canal configuration and similar internal volume were chosen and scanned at an isotropic resolution of 14.16 µm. The sample was assigned to 3 groups (n=10) according to the system used for root canal preparation: Reciproc, WaveOne, and BioRaCe groups. Second and third scans were taken after the canals were prepared with instruments sizes 25 and 40, respectively. The recorded images of the surface area voxels of the canals, before and after preparation were examined from the furcation level to the apex to quantify the non-instrumented surface. Statistical data were compared using GLM for repeated-measures with a significance level set at 5%. Instrumentation systems did not influence the percentage of untouched root canal surfaces (p=0.690) whilst a significant reduction in the percentage of static voxels was observed after the enlargement of the root canal (p=0.010) in all groups (p=0.507). None of the systems was able to prepare the entire surface area of the mesial root canal of mandibular molars. The increased final apical size resulted in a significant positive effect on the shaping ability of the tested systems.


Resumo O objetivo deste estudo foi comparar a porcentagem de área não-instrumentada de canais radiculares preparados com diferentes ampliações utilizando sistemas reciprocantes de lima única (Reciproc e WaveOne) e um sistema rotatório convencional de múltiplas limas (BioRaCe) usando a análise da micro tomografia computadorizada. Trinta raízes mesiais de molares inferiores com curvatura moderada (10º a 20º) apresentando a configuração classe II de Vertucci e semelhança de volume do canal foram selecionadas e escaneadas em uma resolução isotrópica de 14,16 µm. A amostra foi dividida em 3 grupos (n=10) de acordo com o sistema utilizado para a preparação do canal radicular: grupo Reciproc, grupo WaveOne e grupo BioRaCe. Segundo e terceiro escaneamentos foram realizados após os canais serem preparados com instrumentos tamanhos 25 e 40, respectivamente. As imagens registradas de voxels da área de superfície dos canais, antes e após o preparo, foram examinadas desde o nível da furca até o ápice para quantificar a superfície não instrumentada. Os dados foram comparados estatisticamente através de GLM para medidas repetidas com um nível de significância de 5%. Os sistemas de instrumentação não influenciaram o percentual de superfícies não tocadas dos canais radiculares (p=0,690), enquanto que uma redução significativa no percentual de voxels estáticos foi observada após o alargamento do canal radicular (p=0,010) em todos os grupos (p=0,507). Nenhum dos sistemas foi capaz de preparar toda a área de superfície do canal mesial dos molares inferiores. O aumento do tamanho final apical resultou num efeito positivo significativo sobre a capacidade de modelar dos sistemas testados.


Subject(s)
Humans , Alloys , Root Canal Preparation , X-Ray Microtomography/methods